A 64×64 spad array for portable colorimetric sensing, fluorescence and x-ray imaging
Journal article   Peer reviewed

A 64×64 spad array for portable colorimetric sensing, fluorescence and x-ray imaging

C Accarino, G Melino, Valerio Francesco Annese, MA Al-Rawhani, YD Shah, D Maneuski, C Giagkoulovits, JP Grant, S Mitra, C Buttar, …
IEEE Sensors Journal, Vol.19(17), pp.7319-7327
19
2019
Handle:
https://hdl.handle.net/10863/53898

Abstract

We present the design and application of a 64 imes 64 pixel SPAD array to portable colorimetric sensing, and fluorescence and x-ray imaging. The device was fabricated on an unmodified 180 nm CMOS process and is based on a square p+/n active junction SPAD geometry suitable for detecting green fluorescence emission. The stand-alone SPAD shows a photodetection probability greater than 60% at 5 V excess bias, with a dark count rate of less than 4 cps/ μ extm^{\mathbf {2}}} and sub-ns timing jitter performance. It has a global shutter with an in-pixel 8-bit counter; four 5-bit decoders and two 64-to-1 multiplexer blocks allow the data to be read-out. The array of sensors was able to detect fluorescence from a fluorescein isothiocyanate (FITC) solution down to a concentration of 900 pM with an SNR of 9.8 dB. A colorimetric assay was performed on top of the sensor array with a limit of quantification of 3.1~μ extm. X-rays images, using energies ranging from 10 kVp to 100 kVp, of a lead grating mask were acquired without using a scintillation crystal. © 2001-2012 IEEE.
url
https://doi.org/10.1109/JSEN.2019.2916424View